-
Notifications
You must be signed in to change notification settings - Fork 10
07. Filtering Example
This example uses the Normalizer Filter.
When detector deadtime is very high, the detector cannot keep up with
the incoming x-rays. For example, low energy x-rays such as those coming
from Ar
may not get counted and mapping the distribution of Ar
will reveal areas of next to no intensity.
The Normalizer Filer
scales each spectrum so that the intensity at a
given channel is always the same. This filter can be used in this case
to correct the spectra. By normalizing against a channel from within the
Ar
peak, we can mitigate the distortion caused by detector
saturation.
Go to the Filters
tab and add the Normalizer Filter
found under the
Advanced
filter types. Click on the Settings
button to bring
up the settings pane for the filter. Place the cursor on the Ar
peak in the spectrum and read the channel number and value (counts) from
the information found below the spectrum window, and enter them into the
appropriate settings fields. Make sure you are viewing the Mean Average
spectrum to get the average value (counts) for the Ar
.
Close the edit window.
Uncorrected `Ar` with weak signal where `Ti` is strongest
Uncorrected `Ti`
Corrected `Ar`
Corrected `Ti` with greater maximum intensity